PM5 Gen2 Technical Parameters

 Working Language  P-E (Pulse Echo Method): Measures the time between the initial pulse and the first bottom echo.
 I-E (Interface Wave Echo Method): Measures the time between the delayed block echo and the first bottom echo.
 I-E-E (Echo Echo Method): Measures the time between two consecutive bottom echoes.
 I-ME (Multiple Echo Method): Measures the time between multiple bottom echoes.
 E-EV (Double Echo Verification): Verifies the echo pattern, requiring three reflected echoes.
 COAT Coating Mode (Option): Measurement mode specifically for coatings.
 OXIDE Oxide Layer Mode (Option): Dedicated measurement of the thickness of the oxide layer on the inner wall of boiler tubes.
 PEL R Large Range Mode (Option): This mode should be selected when the range exceeds the material thickness.
 Probe Type  Supports delay line probes, contact probes, protective film probes, pen probes, and immersion probes.
 Calibration Method  Zero calibration probe, zero calibration coating, speed of sound calibration, single point calibration, two-point calibration, set speed of sound, material speed of sound table.
 Display Error  ±0.01
 H≤3mm ±0.05                      
 3<H<10mm ±(0.5%H+0.01)    
 H≥10mm               
 Note: H is the thickness of the measured material.
 Thickness Range  0.01mm to 38m, depending on the measured material, probe, surface conditions, temperature, and selected configuration. 0.15mm to 30mm (standard D15A probe, steel)
 Material Speed of Sound Range  310 - 18699m/s, and can reverse measure the speed of sound for known thickness objects.
 Resolution  Low Resolution: 0.1mm Standard Resolution: 0.01mm High Resolution: 0.001mm Ultra High Resolution (Option): 0.0001mm or 0.1um (I-ME mode)
 Probe Frequency Range  0.5MHz to 20MHz

 

General Specification

 Measurement Mode  Available in Chinese and English.
 Measurement Unit  Millimeters, inches, micrometers, mils.
 Measurement Update Rate  4Hz, 8Hz, 16Hz available.
 Instrument Shutdown  Can be set to automatically shut down after 5, 10, or 20 minutes of inactivity, or can only be shut down manually.
 Temperature Range  –10℃ to 50℃
 Instrument Shell  Both the shell and connectors are waterproof sealed, meeting IP67 waterproof and dustproof standards.
 Size  153mm×76mm×37mm (H×W×D)
 Weight  280g including battery.
 Working Power Supply  Two AA batteries, standby time greater than 24 hours.

 

Display

 Display Screen  2.4 inch (320×240 pixel) TFT color LCD screen.
 Rectification Mode  Radio frequency, phase-inverted radio frequency, full wave, positive half wave, negative half wave.
 View Mode  A-scan, B-scan, numerical, maximum value capture, differential.

 

Built-in Data Logger Option

 Data Storage  Identifies, stores, retrieves, clears, and transmits thickness readings, waveform images, and instrument configurations through a standard USB port.
 Capacity  Can store 100,000 thickness values, 400 parameter configuration files, equipped with upper computer software and USB flash drive.
 Real-time Communication Option  Upper computer controls the instrument, real-time transmission of thickness values, waveforms, parameters, including upper computer measurement software. Suitable for online measurement and secondary development.
 Multi-channel Option  Includes upper computer multi-channel measurement software, mainly used for automated thickness measurement.

Ultrasonic Thickness Gauge,Coating Measurement,High Temperature Measurement,Oxide scale measurement